UNI-T UT3000 series Istruzioni per gli Interventi Pagina 34

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 98
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 33
4. Calibration
page 34
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
Waiting time 10 sec.
End of Calibration
If the last calibration standard has been injected, press F1- End. Now the calibration function is
calculated automatically. An outlier is marked as a square. It can be accepted to be included in
the calibration function (F1- Yes) or it can be excluded from the calibration function
(F2- No).
If the outlier is not accepted (F2- No), the user is asked to repeat the calibration point.
On a second page calibration data are
displayed.
Vedere la pagina 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 97 98

Commenti su questo manuale

Nessun commento