UNI-T UT3000 series Istruzioni per gli Interventi Pagina 67

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 98
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 66
6. Maintenance
page 67
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
6.4 Troubleshooting
6.4.1 Unusual Measuring Peaks
An example for a non-typical peak is shown in the picture below.
The peak is more of triangular shape and the signal drops to negative values after the peak.
Figure: Example for an non-typical desorption peak
The reason may be precipitation in the tubing at the outlet of the GoldTrap (= tubing which
connects GoldTrap and optical cell, see picture below).
In this case it is recommended to replace the contaminated tubing.
Figure: Tubing at outlet of GoldTrap. The arrow indicates the
position where precipitation might occur.
Vedere la pagina 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 97 98

Commenti su questo manuale

Nessun commento